Description
HDYSP-H Oil Dissolved Gas Analyzer
Introduction
HDYSP-H adopts the isolated and analysis techniques of multicomponent mixture. It mainly makes use of the differences in the samples of the component boiling point, polarity and the adsorption coefficient in the chromatographic column, to make the components isolated in the chromatographic column, and qualitative and quantitative analyze the isolated components.
HDYSP-H oil chromatographic analyzer in accordance with the power system, "insulating oil dissolved gas component content mensuration (chromatography)" requirement, uses micro-computer control, full keyboard, large screen LCD display, with high precision, stable and reliable performance, high sensitivity, good repeatability etc.
The instrument uses two columns in parallel shunt system. Equipped with TCD and double FID and methane reformer, one sample injection can achieve full analysis of seven components of the oil dissolved gas (H2, CH4, C2H2, C2H4, C2H6, CO, CO2), if necessary can do full analysis of nine components (H2, O2, N2, CH4, C2H2, C2H4, C2H6, CO, CO2)
Features
1.Using advanced technology 10 / 100M adaptive Ethernet communication interface, and built-in IP protocol stack, can easily through the enterprise intranet and Internet for remote data transmission; Convenient for laboratory construction, simplify the laboratory configuration, convenient for management of data analysis
- The Internal of instrument designed three separated connection process, can connect to the local processing site (laboratory), unit director (such as quality inspection chief, production factory director, etc.), and higher authorities (e.g., environmental protection bureau, bureau of technical supervision, etc.). Can easy to real-time monitor the operation of the instrument and the analysis of data results
3.Optional NetChromTM workstation can simultaneously support multiple chromatographic analyzer work, data processing and control, simplifies document management, and reduce the user's investment and operating costs
4.System with Chinese and English operating system, the user can switch according to their needs
5.Temperature control area can be freely named by the user, convenient for use.
6.The instrument uses the multi-processor parallel work mode, stable and reliable; can meet the complex sample analysis, can optional high-performance detectors, such as FID, TCD, ECD, FPD and NPD, can be installed four detectors at the same time.
7.Adopts the modular structure, design and clear, easy to replace and upgrade, to protect the effectiveness of investments
8.The new microcomputer temperature control system, high precision temperature control, superior reliability and anti-jamming performance; with six fully independent temperature control system can be realized 16 stages temperature programming, so the device is qualified for a wider range of sample analysis; with oven automatic back opening door system, the low temperature control accuracy is improved, up / cool down faster
9.Can optional advanced electronic flow controller (EFC), electronic pressure controller (EPC), to realize the digital control, can greatly improve the reproducibility of qualitative and quantitative results
10.with timing self-starting program, it can easily complete on-line analysis of gas samples (equipped with on-line automatic sample parts required)
11.Full computer control keyboard operating system, simple operation, and design a detector with automatic identification technology. With fault diagnosis, and power-off data protection function, automatic memory setting parameters
12.Built-in low noise, high-resolution 24 AD circuit, and with the function of baseline storage and baseline deduction
13.Chromatographic signals acquisition and data processing, suitable for WinXP, Win2000, Windows7 operating systems. In line with the A / A (USA Analysis Society) standard CDF file is read into the sampling data, it can be integrated with Agilent, Waters and other chromatography workstation
14.The chromatographic system with completely independent intellectual property rights, with the MODBUS/TCP standard interface, easy for docking DCS
15.The device can be docked with many of automatic samplers; such as Shimadzu AOC-20i, Italy HTA company's HD series high-performance liquid, gas auto-sampler, etc.
Parameters
Main parameters
Display | 192×64 LCD |
Temperature control area | 6 loops |
Temperature control range | Room temperature: +5℃~400℃, increment: 1℃, accuracy: ±0.1℃ |
Programmed temperature order | 16 |
Program rise rate | 0.1~40℃/min |
Gas path control | Mechanical valve control mode, electronic pressure flow control mode |
external | 4 loops |
Sample injector type | Filling column, capillary tube, six way valve gas inlet, automatic headspace sampling, etc. |
Number of detectors | 5: FID, TCD, ECD, FPD, NPD (optional) |
Start sampling | Manual, auto |
Communication | IEEE802.3 |
Detector parameters
FID | |
Range | ≤3×10-12g/s (n-hexadecane/ isooctane) |
Baseline noise | ≤5×10-14A |
Baseline drift | ≤1×10-13A/30min |
Linear range | ≥107 |
TCD | |
Sensitivity | S≥10000mV.ml/mg(benzene, toluene) (Amplification 1, 2, 4, 8 times optional) |
Number of detectors | 5: FID, TCD, ECD, FPD, NPD (optional) |
Baseline noise | ≤20μV |
Baseline drift | ≤30μV/30min |
Linear range | ≥105 |
ECD | |
Range | ≤1×10-14g/ml(lindane / isooctane) |
Baseline noise | ≤0.03mV |
Baseline drift | ≤0.2mV/30min |
Linear range | ≥104 |
Radioactive Source | 63Ni |
Detection limit | (S)≤2×10-11g/s (P)≤1×10-12g/s (parathion-methyl / Absolute alcohol) |
Baseline noise | ≤3×10-13A |
Baseline drift | ≤2×10-12A/30min |
Linear range | sulfur≥103, phosphorus≥104 |
FAQ:
1. When can I get the price?
We usually quote within 6 hours work time after we get your inquiry. If you are very urgent to get price, please call us or tell us in your email so that we will regard your inquiry priority.
2. What about the delivery time for production?
Honestly, it depends on the order quantity and the product. Usually our leading time is within 7 working days after order confirmation.
3. What is your payment terms?
We accept EXW, FOB, CFR, CIF, DDU, ect. You can choose the one which is the most convenient and cost effective for you.
4. How much will the Shipping cost and time be?
This will depend on size of your goods and method of shipping (by sea/ by air/by Express) and your designated seaport or airport. We will be grateful if you can spare us some minutes to evaluate the cost based on the information of product.
5. Could you please tell me the Warranty Period of your product?
Our warranty period is one year free and lifelong maintenance. If the product is damaged (if not man-made damage), in the first year,we will send free parts for replacement. Over warranty, maintenance is paid service.
6. Do you have Technical Backup to all customers from different countries?
To be sure,our technical support is always available and quick response.
We are manufacturer with much on-site testing experience, thus we know how to deal with your problems. Moreover, our engineers can go to your country to support if needed.
Reviews
There are no reviews yet.