Tech: X Ray Machine Safety Cautions
High voltage danger, watch out for electricity
Ionizing radiation, careful radiation
Prompt attention, safety regulation operation
Illegal operation and danger
The personal protective equipment required is required by the user. All safety equipment
reguires strict and effective regular inspection and maintenance.
Related Product, Video, News, Project,please kindly click the letters to view: |
Product: Overhead Cable/Line Testing Instrument |
VIDEO: Cable/Line/Pipeline Testing Instrument |
NEWS: Cable/Line/Pipeline Testing Instrument |
PROJECT: Cable/Line/Pipeline Testing Instrument |
Please ensure that the controller used is consistent with the generator type and
number, Due to the different attributes of the X ray tube, the incorrect match will damage
the performance and service life of the product.
Please make sure the place used in voltage stability, voltage instability cause high
damage to X ray tube of generator, and easy to upgrade the original controller damage
rate, we ensure the best performance of the original and the highest stability, but the
loss caused by voltage instability is unavoidable.
X ray flaw detector is forced to shut down under normal operation condition,which
will cause great damage to X Ray flaw detector. It is necessary to make the X ray
machine work according to 1:1's working : rest way. Although irregularities will save you
time for testing temporarily, it will cause great damage and lead the equipment in
possible demand for repair.
Every day before use X ray detector, please take time for training, The training time
should be selected according to the length of time the machine is out of service, generally,
the training time to the highest KV should not be less than 5 minutes. When the equipment
is out of service for more than a week, complete machine operation training is necessary.
Proper training will effectively ensure the service life and performance of your X-ray
detector.
Before use, please ensure that the insulation pressure of the X-ray flaw detector is
not less than the minimum value of 0.45MPa. If it is found to be lower than this value or
other unexpected failures, please stop using it and contact us to ensure that the losses
caused by X-ray flaw detector damage are minimized.
The series X ray flaw detectors shall be operated by trained professionals.
Maintenance and repair is responsible for the designated professionals who know the
risks and risks of the system.
Without the permission of the manufacturing company, any non designated
maintenance personnel shall not dismantle the equipment shell. Without the written
approval of manufacturing company, it is prohibited to change, expand, or replace the
structure of the system.
The owner of the system is responsible for the use of the system and compliance
with the regulations of the law.
This system is operated by people who are prohibited from lack of technical ability.
It is prohibited to ignore, remove or change the safety devices that the system should be
equipped with.
Without the permission of the manufacturing company, any non designated
maintenance personnel shall not dismantle the equipment shell. Without the written
approval of manufacturing company, it is prohibited to change, expand, or replace the
structure of the system.
When the X ray flaw detector opens high pressure, it produces X rays, which can
cause harm to health. Therefore, we must obey every warning so as not to be exposed
directly to the X ray radiation area.
1.the direction of X rays should be directed toward the earth or wall (preferably the
wall).
2.a thick lead screen is provided behind the samples to be checked in order to absorb
all radiation. For example, 160KV uscs 4mm lead, 200KV for 5mm lead, 300KV for 12mm
lead, this is just for your reference, actural thickness is depended on your case.
3.X rays are used in the surrounding area to allow the number of pcople to be reduced
as much as possible.
4.indoor use should be equipped with protective lead room, or special laboratory.
5.distance protection may be dangerous even at the distance of 100 meters of the flaw
detector. It is suggested that the operator should be equipped with a ray warning device
to better protect the radiation.
The X ray detector is made up of a circuit to produce high voltage.
Please connect the ground wire to the ground reliably. Pay attention to the danger of high
voltage electricity. Only the designated professional can dismantle the cover of the
equipment. Only designated professionals can do external electrical connections and
apply power to the equipment.
When the power is energized, no person is allowed to do any action against the safe
operation of the high voltage circuit so as not to cause an unexpected personal damage.
Portable X ray detector should be stored in a dry, clean, far away heat source and
non direct sunlight.
The total power supply should be cut off when cleaning the portable X ray detector.
It should be ensured that no liquid penetrated into the equipment to avoid electrical
circuit short circuit and corrosion of the internal components.
It is strictly prohibited to use corrosiveness, solubility, abrasiveness detergent or
grinding agent to clean. The painted parts and the outer surfaces of the aluminum parts
can be cleaned with wet cloth and soft cleaners, andthen dry with dry cotton.
Please keep the cable socket and the plug part clean. In the storage process of the
equipment, the storage is closed with a separate layer.
Portable X ray flaw detection of non voltage energized state will not cause harm to
users and environment.
Electrical waste can not be treated with the usual industrial or household waste.
Please deal with it in an environmentally friendly way.
After connecting all cables, the necessary inspection should be made on the portable
X ray detector and the use environment before the use of the ma-chine is reopened.
Check whether the X ray detector generator barometer is less than 0.45Mpa,if less
than 0.35 Mpa, please stop using it immediately, so as not to cause damage.
Whether the voltage of the inspection field is stable and the unstable input voltage will
cause a lot of loss to the portable X ray detector.
The failure to rest in accordance with the 1: 1 work will cause great damage to the
equipment.